DOI:

Received November 28, 1988,Revised , Accepted , Available online

Volume 1,1989,Pages 104-108

  • Summary
  • References
  • Related Articles
A direct and simple method for the clinic muti-element determination of human blood sample by Zeeman graphite furnace atomic absorption spectropho-tometry is reported. No pretreatment of the sample is necessary other than simple dilution of the sample with aqueous solution of Triton X-100. The detection limits of the elements investigated fall in the range of 0.1- 4.4 μg/L with a relative standard deviation ranging from 1.6 for zinc to 3.73% for lead. Micro liters of samples are sufficient for the determination of each element.

Copyright © Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences. Published by Elsevier B.V. and Science Press. All rights reserved.京ICP备05002858号-3