DOI:

Received December 31, 1997,Revised May 13, 1998, Accepted , Available online

Volume 11,1999,Pages 48-56

  • Summary
  • References
  • Related Articles
The GC determination of aldehydes and ketones in stack gas was reported. The stack gas was collected with the dust sampler by isokinetic sampling technique, 2,4 dinitrophenylhydrazine in acidic solution was used to absorb aldehydes and ketones in stack gas specifically, to form their hydrazones, which were determined by GC with a capillary column of SE 54. The GC system can separate nine hydrazones of carbonyls interested with constant recovery rates of 94%—97%, giving lowest detectable limits of 0 05—1 0 ng. The standard gas of carbonyls were determined by this method with relative errors of -5% to -10%, percent coefficients of variation within 5%, and spiking recovery rates of 93%—97%. As an application of the method, the carbonyl pollutants were determined in stack gas samples of a chemical plant.