A rapid, simple and sensitive method was demonstrated for the determination of phenolic compounds in water samples by alternatingcurrent oscillopolarographic titration. With the presence of sulfuric acid, phenol could be transferred into a nitroso-compound by reacting with NaNO2. The titration end-point was obtained by the formation of a sharp cut in the oscillopolarographic with infinitesimal NaNO2 on double platinum electrodes. The results showed that phenol concentration had an excellent linear relationship over the range of 4.82 10??6–9.65 10??3 mol/L, the RSD of the proposed method was lower than 1.5%, and the spiked recoveries of three real water samples were in the range of 95.6%–106.9%.

Copyright © Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences. Published by Elsevier B.V. and Science Press. All rights reserved.京ICP备05002858号-3